| Treatment Level Range | 0.22 SMT to 1.6 SMT (calibrated linear output) |
| Induced Electric Field at 2cm at 1.0 SMT | 135 V/m (nominal) |
| %MT Range | 80% to 140%MT |
| Pulses Per Second (pps) Range | 0.1-30 pps (also known as Hz) |
| Stimulation Time (pulse train duration) Range | 1-20 seconds for > 1 pps 1-30 seconds for 1 pps |
| Inter-Train Interval Range | 0-60 seconds for 1 pps 10-60 seconds for > 1 pps |
| Pulse Type | Biphasic sinusoid |
| Pulse Width | 185µS (nominal) |
| Maximum Pulses per Treatment Session | 5000 |
| Programmable Pulse Sequences | Ten custom pulse sequences plus the default standard sequence |
| Coil Type | High efficiency, low temperature, field-focusing ferromagnetic core |
| Coil Positioning System | Integrated into head support, Accurate and repeatable laser aided coil placement |
| Treatment Quality Features | Field level detection, Coil contact sensing |
| MT Level Determination Method | Proprietary MT Assist® algorithm |
| Patient Comfort Features | Surface field reduction, Reclining chair & head support |
| User Controls | 15" Color LCD Touchscreen Graphical User Interface |
| Patient Data Security | HIPAA compliance, Back-up and archive |
| Safety | Certified to UL60601-1 & IEC 60825-1 NINDS limits built-in |